Vacheck

Vacheck
Easy to Use
  • Fits any 1/2" HVE valve
  • Simple "Pass-Fail" indication
  • Fixed calibration, no adjustments
Evaluates Dental Vacuum Suction
  • Use in Flowing HVE valve
  • Flow tests vacuum station
  • Detects clogged lines
PRODUCT NO. 800500
Patent 5,779,472


RAMVAC VACHECK Uses

Check treatment room for minimal useful vacuum performance. Ensure dental vacuum system can support safe, efficient operation of high volume evacuators and nitrous oxide scavengers.

Specify size for proposed systems

Example, specify: The vacuum system must provide acceptable perfomance for # [specified number of] HVE equivalents (1 HVE equivalent = 1 HVE or 5 saliva ejectors or 5 small surgical aspirators or 4 nitrous oxide scavengers) in operation simultaneously and continuously, as demonstrated by a RAMVAC Vacheck "pass" indication.

Acceptance check - New equipment

Example, specify:Before accepted, the vacuum system must provide an acceptable performance for # [specified number of] HVE equivalents (1 HVE equivalent = 1 HVE or 5 saliva ejectors or 5 small surgical aspirators or 4 nitrous oxide scavengers) in operation simultaneously and continuously, as demonstrated by RAMVAC Vacheck "pass" indication.  Performance check - Old equipment Repeated periodic Vacheck testing can identify clogged lines and poorly cleaned solids separators and vacuum hoses.

Accurate
  • Responds to small changes in flow or vacuum strength
  • High repeatability

Construction
  • Stainless steel flow bushing
  • PVC flow tube
  • Precision machined to ten thousandths of an inch